The ATT1000 Series two-wire input temperature transmitter is a high performance device that receives Thermocouple / RTD / ohm / mV input and output mA with HART communication protocol.Three internal keys to set parameter. Zero span calibration can be achieved either by HART or keypad.

SPECIFICATION
Inputs:Display:
Universal InputBacklit LCD rotatable to any angle
ThermalcoupleDisplay Function:
T/C (K, S, B, E, J, T, R, N)……5 Digits programmable & 0-100% Bargraph
RTDDisplay Unit:
CU50 / PT100, 2 / 3 / 4 wire……oC / oF / oR / oK
mVSetting Method:
-128~128 mV / -1000~1000 mV……B Type
ResistanceUSB + PC software (windows system)
0-600 / 0-5000C Type
Accuracy:HART signal
+/-0.1% of spanD Type
Stability:HART signal & keypad
+/-0.1% of span for 2 yearsCurrent Output:
Temperature Effect:4-20 mA 2 wires with HART Signal
+/-0.1% of span per 10 oCLoad
CJC Compensation:Rohm = (VDC-9) * 50
+/-0.5 oCPower Supply:
Simulation Output:9-36 VDC
4-20 mADigital Communication:
Material:HART Protocol
Wetted PartsCable Entry:
S.S. 304 / S.S. 316 / Has. C / Tantalum etc.M20 Conduit Threads / 1/2″ NPT (Female)
HousingDamping:
Low copper aluminum alloy (Sprayed with0-32 Seconds
light blue polyurethane paint)Response Time:
Module Housing100 ms
Flame-resistant PCTurn on Time:
Name / Tag Plate2 Seconds with minimum damping
Stainless Steel 304 / Stainless Steel 316Zero Calibration:
Process Connection:Automatic zero calibration by press-button
Threads / FlangesIsolation:
Over Temp. Limit:Input / Output / GND isolated to 480 VDC
120% of Full ScalePower Supply Effect:
Mounting:+/-0.005 of Span per V
Direct mounting or bracket on 2″ PipeWeight:
Humidity Limit:60 g (Converter only) with display
0-100% Relative Humidity30 g (Converter only) without display
Ambient Temperature:Protection Class:
-25~85 oCIP67 (Standard)
Vibration Effect:Intrinsically Safe, Eex ia IIC T5 (Standard)
+/-0.05% of URL per g to 200 Hz in anyExplosion proof, Ex d IIB T5
axisSensor Connection:
EMI / RFI Effect:M27 * 2 (Female)
Follow SAMA PMC 33.1 from 20-1000
MHz and for field strengths up to 30 V/m